Z-0528-2013Class II

Mindray V12 Size: 320 mm X 320 mm X 450 mm N.W.: 3 kg G.W.: 6 kg Qty:1 Manufactured in China Mindray V21...

Datascope Corp. / initiated 2012-09-24 / Terminated / root cause: manufacturing process

Mindray DS USA, Inc. d.b.a. Mindray North America began a recall of Mindray V12 Size: 320 mm X 320 mm X 450 mm N.W.: 3 kg G.W.: 6 kg Qty:1 Manufactured in China Mindray V21 Size: 515 mm X 335 mm X 685 mm N.W.: 8 kg G.W.: 11 kg... on . FDA classified it as Class II and gives the reason as "Mindray has identified an issue with the V Series Monitor where the monitors touch screen may stop responding to touch". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0528-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Mindray V12 Size: 320 mm X 320 mm X 450 mm N.W.: 3 kg G.W.: 6 kg Qty:1 Manufactured in China Mindray V21 Size: 515 mm X 335 mm X 685 mm N.W.: 8 kg G.W.: 11 kg Qty:1 Manufactured in China One product, V Series monitor, available in two sizes: The V12 has a 12 inch screen, the V21 has a 21 inch screen. The V Series Monitor is a Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices. It also monitors of the following human physiological parameters: 1) ECG waveform derived from 3, 5, 6 and 12 lead measurements, 2) Heart Rate, 3) Pulse Oximetry (Sp O2), 4) ST Segment Analysis, 5) Arrhythmia Detection, 6) Non Invasive Blood Pressure (NIBP), 7) Invasive Blood Pressure (IBP), 8) Cardiac Output (CO), 9) Respiratory Gasses, 10) Respiration Rate, 11) Temperature, It is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Mindray has identified an issue with the V Series Monitor where the monitors touch screen may stop responding to touch.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Mindray issued an "Urgent: V Series Monitoring System Corrective Action" letter dated 9/24/2012 via certified mail with return receipt verification to all hospitals. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact their Mindray Service Representative at 1-800-288-2121 to arrange for relplacement of the affected product. For questions regarding this recall call 201-995-8407.
Quantity in commerce
206 units (180 units domestic, 26 units foreign)
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Great Britian, Colombia, and Australia.
Product codes and lots
V Series Monitor P/N's 0998-00-1800-101 and 0998-00-1800- 201.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0528-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0528-2013Class IITerminatedVoluntary: Firm initiated