DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl
DePuy Spine, Inc. began a recall of DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component on . FDA classified it as Class II and gives the reason as "Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0712-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DePuy Spine, Inc.
- Product
- DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
- FDA root cause
- Component change control
- Action
- DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to [email protected]. For questions call Linda Regan @508-977-6606.
- Quantity in commerce
- 6 units
- Distribution
- Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0705-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0706-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0707-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0708-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0709-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0710-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0711-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0712-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0713-2013 | Class II | Terminated | Voluntary: Firm initiated |