Z-0439-2013Class II

Cooper Surgical RUMI Arch, Reusable Uterine Manipulator Handle Ref: UMH700 The intended use of the RUMI Arch...

CooperSurgical, Inc. / initiated 2012-10-31 / Terminated / root cause: design

Cooper Surgical, Inc. began a recall of Cooper Surgical RUMI Arch, Reusable Uterine Manipulator Handle Ref: UMH700 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a... on . FDA classified it as Class II and gives the reason as "Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0439-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CooperSurgical, Inc.
Product
Cooper Surgical RUMI Arch, Reusable Uterine Manipulator Handle Ref: UMH700 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy
Product code
LKF Cannula, Manipulator/Injector, Uterine
Reason for recall
Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle
Root cause tag
design
FDA root cause
Device Design
Action
CooperSurgical issued notification on October 31, 2012, to all affected customers Fed X . The notification identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return units for replacement. Customers with questions should contact Customer Serivce at 1-203-601-9818 or 800-243-2974. Customers were asked to fill out the attached Replacement of Recalled Product Form and fax it to 1-800-262-0105. For questions regarding this recall call 203-601-9818.
Quantity in commerce
1215 total
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Australia, Belgium, Canada, Czech Republic, Germany, Italy, Spain, France, United Kingdom, Japan, New Zealand, South Korea, Saudi Arabia, Sweden, Singapore, Slovenia, and Taiwan.
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0439-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0439-2013Class IITerminatedVoluntary: Firm initiated
Z-0440-2013Class IITerminatedVoluntary: Firm initiated