Z-0594-2013Class I
Zimmer Spine Ardis Inserter, Rx, non-sterile
Zimmer GmbH / initiated 2012-11-12 / Terminated / root cause: design
Zimmer, Inc. began a recall of Zimmer Spine Ardis Inserter, Rx, non-sterile on . FDA classified it as Class I and gives the reason as "Zimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interbody Spacer breaks during implantation...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0594-2013
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Spine Ardis Inserter, Rx, non-sterile
- Product code
- MAX Intervertebral Fusion Device With Bone Graft, Lumbar
- Reason for recall
- Zimmer Spine has received reports of the PEEK Ardis Interbody Spacer breaking into fragments when too much lateral force is applied to the Ardis Inserter. If the PEEK Ardis Interbody Spacer breaks during implantation, surgical delays of up to 60 minutes could occur. Sharp edges of the broken implant may contact and damage adjacent structures, requiring surgical intervention. Observed health risks
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Zimmer Spine, sent a "MEDICAL DEVICE CORRECTION" memo dated November 12, 2012 to all consignees / customers describing product, problem and actions to be taken including revised surgical technique. Beginning the week of December 17, 2012, Zimmer began the immediate retrieval of the affected units as a Class I recall. December 20, 2012 nation wide press outlining the recall was issued. Customers may report adverse events or quality problems experienced with the use of these products to Zimmer Spine by calling 866-774-6368 between the hours of 8AM and 5PM CST or by email at [email protected].
- Quantity in commerce
- 315 (172 in USA)
- Distribution
- Worldwide distribution: USA (nationwide) including states of: AR, AZ, CA, CO, CT, DC, FL, GA, ID, LA, MD, MI, MN, MO, NC, NV, NY, OH, OK, PA, TN, TX, WI, and WV; and countries of: Australia, El Salvador, France and Singapore.
- Product codes and lots
- Ardis Inserter, Part number: 3256-01, Lot: 52ZW, 50CL, 53BB, 55AB, 58GS, 58GT, 60JY, 60JZ, 62KP, 63HJ, 64EY, 68AX, 71JE, 75LE Part, 07.00780.701, lot 76CN; part 07.00780.702, lot 76CP;part 07.00900.701, lot 76TR; 07.00900.702, 76TS; 650-0297-01, 70YT; 07.00780.701, 73EC; 07.00780.702, 73EC; 07.00900.701, 76TQ; 650-0239-01, 66PR; 650-0293-01, 70SG ; 650-0293-01, 70SG-R; 650-0179-01, 60CK; 650-0176-01, 60CL; 650-0365-01, 72LP; 07.00780.701, 76CQ; 07.00780.702, 76CR; 07.00780.701, 74EN; 07.00780.702, 74EP; 650-0351-01, 72JH; 650-0351-02, 72JJ; 650-0263-01, 68HL; 650-0208-01, 62FH; 07.00900.701, 76TN; 07.00780.701, 76QG; 07.00780.701, 76LC; 07.00780.702, 76LD;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0594-2013 | Class I | Terminated | Voluntary: Firm initiated |