Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0474-2013Class II

Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant)

Zimmer GmbH / initiated 2012-11-01 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS Knee Reamer Sleeves are primarily used to guide the... on . FDA classified it as Class II and gives the reason as "The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0474-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Bio-Modular Humeral Head 48mm x 24mm Ti 6AI4V/Ion Implanted (Orthopedic Implant) Product Usage: The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.
Product code
KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented
Reason for recall
The items in this lot are missing the ion implantation feature. The surface of the head may be softer and more prone to scratching, which in tum could potentially cause higher wear of the poly bearing. Increased wear could potentially lead to a shorter implant life.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Biomet, Inc. sent an URGENT MEDICAL DEVICE RECALL NOTICE dated November 1, 2012 to all affected consignees. The letter identified the affected product, problem and actions to be taken. The notification letter instructed the surgeon to montior the patient as their is a potential for increased wear which have an unknown effect on implant life. For questions call 574-372-1570.
Quantity in commerce
4
Distribution
USA Nationwide Distribution in the state of WI.
Product codes and lots
Catalog Number: 113740 and Lot Number Identification: 779270

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0474-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0474-2013Class IITerminatedVoluntary: Firm initiated