Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0527-2013Class II

EBI XFIX DFS Rail System (Adult Rail Gradual Swivel Clamp). The Biomet Rail System consists of a unilateral...

Zimmer GmbH / initiated 2012-11-09 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of EBI XFIX DFS Rail System (Adult Rail Gradual Swivel Clamp). The Biomet Rail System consists of a unilateral external fixator rail, modular bone screw clamps... on . FDA classified it as Class II and gives the reason as "It was reported that the number scale/graduation markings were missing on 14 pieces of the swivel clamp". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0527-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
EBI XFIX DFS Rail System (Adult Rail Gradual Swivel Clamp). The Biomet Rail System consists of a unilateral external fixator rail, modular bone screw clamps, and compression and distraction units. The device is used in the treatment of bone fractures.
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
It was reported that the number scale/graduation markings were missing on 14 pieces of the swivel clamp.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
On 11/9/2012 an URGENT MEDICAL DEVICE RECALL NOTIFICATION letter was sent to all consignees (distributors and Doctor notices) informing them of the issue and providing recommended actions. Customer contact number (574) 372-1570 Monday through Friday from 8 am to 5 pm.
Quantity in commerce
49
Distribution
Worldwide Distribution, including the states of TX, TN, PA, CA, NY, FL, and MI and the countries of Japan and Columbia.
Product codes and lots
Catalog Number 06350 and Lot Number Identification 994790 (packaged as 042210)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0527-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0527-2013Class IITerminatedVoluntary: Firm initiated