Z-0551-2013Class II

Persona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size D Left; Size D Right...

Zimmer GmbH / initiated 2012-11-14 / Terminated / root cause: design

Zimmer, Inc. began a recall of Persona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size D Left; Size D Right This device is used with the Persona Knee System... on . FDA classified it as Class II and gives the reason as "All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0551-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Persona, The Personalized Knee System, Tibial Sizing Plate, Cemented, Nonsterile, Size D Left; Size D Right This device is used with the Persona Knee System for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening
Root cause tag
design
FDA root cause
Device Design
Action
Zimmer sent distributors beginning the week of November 14, 2012, E-mail information requesting all equipment be collected in distributors office before shipment back to Zimmer and replacement. A telephone conference discussing the recall and amended surgical technique was held the week of November 14, 2012, and Urgent Recall notices were sent to Distributors describing the devices; risks; updated surgical technique with directions to hand deliver surgeon/risk manager letters to all affected surgeons and risk managers in hospitals receiving the product. The notices included directions to remove all affected product and ensure risk manager letter is provided. The letter stated a Zimmer Inc. employee will conduct onsite reconciliation and asked return of distribution Spreadsheet to [email protected] within 24 hours of receiving notification. Product is to be returned with certificate of sterilization to Zimmer Product Service Department 1777 West Center St. Warsaw, IN 46580 or Zimmer International Logistics, Eschbach Germany for International Accounts. For questions regarding this recall call 574-372-4807.
Quantity in commerce
189
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.
Product codes and lots
Part Number 42-5399-067-01(left) Lot 62050138, 62121496  Part Number 42-5399-067-02 (right), Lot 62067442, 62075280, 62121527

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0551-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0539-2013Class IITerminatedVoluntary: Firm initiated
Z-0540-2013Class IITerminatedVoluntary: Firm initiated
Z-0541-2013Class IITerminatedVoluntary: Firm initiated
Z-0542-2013Class IITerminatedVoluntary: Firm initiated
Z-0543-2013Class IITerminatedVoluntary: Firm initiated
Z-0544-2013Class IITerminatedVoluntary: Firm initiated
Z-0545-2013Class IITerminatedVoluntary: Firm initiated
Z-0546-2013Class IITerminatedVoluntary: Firm initiated
Z-0547-2013Class IITerminatedVoluntary: Firm initiated
Z-0548-2013Class IITerminatedVoluntary: Firm initiated
Z-0549-2013Class IITerminatedVoluntary: Firm initiated
Z-0550-2013Class IITerminatedVoluntary: Firm initiated
Z-0551-2013Class IITerminatedVoluntary: Firm initiated
Z-0552-2013Class IITerminatedVoluntary: Firm initiated
Z-0553-2013Class IITerminatedVoluntary: Firm initiated
Z-0554-2013Class IITerminatedVoluntary: Firm initiated
Z-0555-2013Class IITerminatedVoluntary: Firm initiated