Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0570-2013Class II

BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton...

Becton Dickinson & Co / initiated 2012-11-06 / Terminated / root cause: manufacturing process

Becton Dickinson & Co. began a recall of BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, USA. The BD Phoenix" Automated... on . FDA classified it as Class II and gives the reason as "An antibiotic susceptibility test kit may produce false intermediate or false resistant results with clinically significant bacteria". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0570-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, USA. The BD Phoenix" Automated Microbiology system is intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
An antibiotic susceptibility test kit may produce false intermediate or false resistant results with clinically significant bacteria.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, BD, sent an "URGENT PRODUCT RECALL" letter dated November 2012 via the telephone and fax to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and discard any remaining packages for replacement and to complete and return the Acknowledgement Receipt Form, whether or not you have any inventory remaining, via fax to: BD Regulatory Compliance at 410-316-4258. BD will issue replacements for the discarded materials. If further assistance is needed regarding replacements please contact BD Customer Service Department at 1-800-675-0908. For all other inquiries, please contact BD Technical Services Department at 1-800-638-8663.
Quantity in commerce
12,250 tests
Distribution
Nationwide distribution: USA including states of: AR, CA, KY, OK, RI, WI, and WY.
Product codes and lots
Lot number 2053373

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0570-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0570-2013Class IITerminatedVoluntary: Firm initiated