BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton...
Becton Dickinson & Co. began a recall of BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, USA. The BD Phoenix" Automated... on . FDA classified it as Class II and gives the reason as "An antibiotic susceptibility test kit may produce false intermediate or false resistant results with clinically significant bacteria". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0570-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, USA. The BD Phoenix" Automated Microbiology system is intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria.
- Product code
- LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
- Reason for recall
- An antibiotic susceptibility test kit may produce false intermediate or false resistant results with clinically significant bacteria.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, BD, sent an "URGENT PRODUCT RECALL" letter dated November 2012 via the telephone and fax to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and discard any remaining packages for replacement and to complete and return the Acknowledgement Receipt Form, whether or not you have any inventory remaining, via fax to: BD Regulatory Compliance at 410-316-4258. BD will issue replacements for the discarded materials. If further assistance is needed regarding replacements please contact BD Customer Service Department at 1-800-675-0908. For all other inquiries, please contact BD Technical Services Department at 1-800-638-8663.
- Quantity in commerce
- 12,250 tests
- Distribution
- Nationwide distribution: USA including states of: AR, CA, KY, OK, RI, WI, and WY.
- Product codes and lots
- Lot number 2053373
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0570-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY