Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0556-2013Class II

Invacare Height Adjustable (walking) Canes, Model Numbers: 8916, 8917, and 40918-6. The device is packaged...

Invacare GmbH / initiated 2012-11-06 / Terminated / root cause: design

Invacare Corporation began a recall of Invacare Height Adjustable (walking) Canes, Model Numbers: 8916, 8917, and 40918-6. The device is packaged six (6) canes per shipping carton. The cane is used... on . FDA classified it as Class II and gives the reason as "Following their receipt of several customer complaints, Invacare recalled their height adjustable walking canes. The device was recalled from distribution based on reports indicating that there is a possibility that the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0556-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Invacare GmbH
Product
Invacare Height Adjustable (walking) Canes, Model Numbers: 8916, 8917, and 40918-6. The device is packaged six (6) canes per shipping carton. The cane is used as an aid in walking or ambulation.
Product code
KHY Cane, Safety Walk
Reason for recall
Following their receipt of several customer complaints, Invacare recalled their height adjustable walking canes. The device was recalled from distribution based on reports indicating that there is a possibility that the button which holds the cane at the desired height, may suddenly and unexpectedly compress, causing the cane to collapse and cause possible injury to the user.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Invacare, sent an "URGENT MEDICAL DEVICE RECALL (Removal)" letter dated November 6, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to examine their existing stock using the enclosed detailed 'Product Tracking Sheet' and quarantine any affected canes; Acknowledge receipt and understanding of the urgent recall notice and indicate the number of canes remaining in inventory on the enclosed, 'Provider Response Card;'. Fax the completed Provider Response Card to Invacare Regulatory Affairs at: 1- 440-326-3544; disassemble each cane and discard both pieces in the trash; (Disassembly instructions are also provided); and lastly, contact Invacare Customer Service at 1-800-668-2337, Monday - Friday, 8 a.m. - 8 p.m. EST, to order replacement merchandise at no charge. If you have any questions concerning these instructions, contact Invacare Customer Service at 1-800-668-2337, Monday - Friday, 8 a.m. - 6 p.m. ET.
Quantity in commerce
38,280 canes
Distribution
Worldwide distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN KS, KY, LA,MA, MD ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH,OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and countries of: Australia, Puerto Rico, and Taiwan Province of China.
Product codes and lots
The Lot Codes subject to recall are: PW111201, PW120201, PW120101, PW120301, and PW120401

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0556-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0556-2013Class IITerminatedVoluntary: Firm initiated