Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0533-2013Class II

Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1

Zimmer GmbH / initiated 2012-11-15 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1 Product Usage: Knee joint replacement prostheses intended for application with or without... on . FDA classified it as Class II and gives the reason as "Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in surgery greater than 30...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0533-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet Offset Tibial Tray 2.5 mm Adaptor, REF 141490 Sterile, QTY.1 Product Usage: Knee joint replacement prostheses intended for application with or without bone cement. Knee joint replacement components include femoral, tibial, and patellar components, intended for both primary and revision applications. Indications: 1) painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis when one or more compartments are involved, 2) correction of varus, valgus, or posttraumatic deformity, 3) correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure
Product code
MBV Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uhmwpe, Pegged, Cemented, Polymer/Metal/Polymer
Reason for recall
Biomet investigation identified that the square end of the offset tibial tray adaptor locking insert is oversized and may not engage insert tool screwdriver. This could cause a possible delay in surgery greater than 30 minutes, potentially exposing the patient to the increased risks from being under anesthesia for a longer duration.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Biomet Recall sent an Urgent Medical Device Recall Notices letter dated November 15, 2012 to distributors, hospital operating room(OR)managers and Surgeons receiving the affected product. The letter identified the affected product, the reason for the recall and actions to be take, including directions to immediately locate and discontinue use of the affected product. Customers were asked to confirm receipt of the notice by calling 1-888-348-9500 extension 1570. Customers with affected product to be returned are asked to obtain RGA # by calling 574-372-6677 or email: [email protected]. A FAX Back Response Form is to be filled out and sent within 3 days of receipt to 574-372-1683. Return product to the Attention of Audrey Daenzer, Field Action Coordinator, Biomet, Attention Return Goods, Building B Recalls, 56 East Bell Drive, Warsaw, IN 46580. Questions related to the notice should be directed to 474-372-3983 or 574-371-3009, Monday- Friday 8 a.m. to 5 p.m.
Quantity in commerce
200
Distribution
Worldwide distribution - USA (nationwide) and the countries of Australia, Canada, Chile, Europe, Japan, Korea, Panama, Puerto Rico and Thailand.
Product codes and lots
Catalog Number: 141490 Lot 378330,378320,407230,485010,505140,485000,506070,533400,54048, 439700,797820

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0533-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0533-2013Class IITerminatedVoluntary: Firm initiated