Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0621-2013Class I

Fisher & Paykel Healthcare reusable breathing circuit. Non-heated breathing circuit, 1.8m or 6 feet in...

Fisher & Paykel Healthcare, Inc. / initiated 2012-11-19 / Terminated / root cause: manufacturing process

Fisher & Paykel Healthcare, Ltd. began a recall of Fisher & Paykel Healthcare reusable breathing circuit. Non-heated breathing circuit, 1.8m or 6 feet in length. Oxygen therapy delivery system for adult patient on . FDA classified it as Class I and gives the reason as "FPH has become aware that one batch of tubes supplied to FPH may contain hole damage. These tubes were incorporated by FPH into certain 900MR068 reusable circuits with lot numbers 110810 and 111020. If this hole damage...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0621-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fisher & Paykel Healthcare, Inc.
Product
Fisher & Paykel Healthcare reusable breathing circuit. Non-heated breathing circuit, 1.8m or 6 feet in length. Oxygen therapy delivery system for adult patient.
Product code
BZO Set, Tubing And Support, Ventilator (W Harness)
Reason for recall
FPH has become aware that one batch of tubes supplied to FPH may contain hole damage. These tubes were incorporated by FPH into certain 900MR068 reusable circuits with lot numbers 110810 and 111020. If this hole damage is not detected during the standard leak test before patient use, it could potentially result in a gas leak in the breathing system, which may lead to a loss of pressure for the int
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Fisher & Paykel sent an Urgent Medical Device Recall letter dated November 19, 2012, with a Product Recall Response Form to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to destroy and discard any affected circuits in their possession. Customers were instructed to complete and return the Product Recall Response form to the address indicated on the form, even if they have no affected product. Customers were asked to forward the recall notice to all those who need to be aware within their organization. Customers with questions should call 1-800-792-3912. For questions regarding this recall call 949-443-4095.
Quantity in commerce
200 units total, 26 units US
Distribution
Worldwide Distribution - USA including IL, IN, AR and GA.
Product codes and lots
Lots 110810 and 111020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0621-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0621-2013Class ITerminatedVoluntary: Firm initiated