Z-0742-2013Class II
Mindray DPM5 Monitor, a vital signs monitor used on human patients in healthcare facilities
Datascope Corp. / initiated 2012-09-05 / Terminated / root cause: design
Mindray DS USA, Inc. d.b.a. Mindray North America began a recall of Mindray DPM5 Monitor, a vital signs monitor used on human patients in healthcare facilities on . FDA classified it as Class II and gives the reason as "Mindray has identified an issue with the DPM5 monitors where ST and Arrhythmia features may not be available". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0742-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- Mindray DPM5 Monitor, a vital signs monitor used on human patients in healthcare facilities.
- Product code
- MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
- Reason for recall
- Mindray has identified an issue with the DPM5 monitors where ST and Arrhythmia features may not be available.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Mindray DS USA Inc. sent an Urgent Product Corrective Action letter on September 5, 2012 via Fed Ex with delivery confirmation to all users. The letter identifies the affected product, problem and actions to be taken. Customers were instructed to contact their Mindray Service Representative at 1-800-288-2121 Monday through Friday 8:30-5:30 pm, to schedule a software upgrade.
- Quantity in commerce
- 13 units
- Distribution
- US Nationwide Distribution including the states of: VA, TN, FL, & TX
- Product codes and lots
- p/n 9201F-PA000XX DS- 0C000565 02000335 12000603 12000599 12000597 OCOOO569 9A000247 9A000252 17000648 06000490 05000470 05000466 02000373
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0742-2013 | Class II | Terminated | Voluntary: Firm initiated |