Z-0674-2013Class II

Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Scanning Systems: Brilliance 64 - Model #728231; and...

Philips North America LLC / initiated 2012-12-03 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Scanning Systems: Brilliance 64 - Model #728231; and Ingenuity CT - Model # 728326, Distributed By... on . FDA classified it as Class II and gives the reason as "Software anomalies. Philips Medical Systems recalled their Brilliance 64 running software v4.0.0xx516 and their Ingenuity CT running software v4.0.0xx518 due to software issues detected in these CT scanning systems. If...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0674-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance 64 and Ingenuity CT Computed Tomography X-Ray Scanning Systems: Brilliance 64 - Model #728231; and Ingenuity CT - Model # 728326, Distributed By: Philips Healthcare, Cleveland, OH. The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Software anomalies. Philips Medical Systems recalled their Brilliance 64 running software v4.0.0xx516 and their Ingenuity CT running software v4.0.0xx518 due to software issues detected in these CT scanning systems. If these issues were to re-occur they could pose the risk of injury to patients undergoing scans; device users; or device service personnel.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Philips Healthcare, sent an 'URGENT- Medical Device Correction" letters' dated November 15, 2012, and December 3, 2012, to their customers. Both letters describe the product, problem and actions to be taken. The letters state that Philips is notifying all affected customers of this information via a 'Field Safety Notice and will be installing a software update on the affected Brilliance 64 and Ingenuity CT system in order to address these issues. A Philips' Field Service Engineer will be contacting each customer site and make arrangements to install the necessary software update on these device systems. The customers should contact their local Philips representative or local Philips Healthcare office. In North America and Canada needing further information or support concerning this update, should contact our Customer Care Solutions Center at 1-800-722-9377. Selection option 5, and enter their site ID or follow the prompts.
Quantity in commerce
A total of nine (9) units were distributed (3 units in the U.S & 6 units overseas)
Distribution
Worldwide distribution: USA including IN, OH and OR; and the countries: Australia, Belgium, China, Israel, and Sweden.
Product codes and lots
Brilliance CT X-RAY System - Model/System Code #728231, Serial Number distributed in the U.S. was: 9654. Ingenuity CT X-Ray System - Model/System Code #728326 - Serial Numbers distributed in the U.S. were: 300018, and 300003. The remaining Serial Numbers distributed overseas were: 91003, 95130, 95157, 95504, and Serial Numbers: 300010, and 300030 .

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0674-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0674-2013Class IITerminatedVoluntary: Firm initiated