Z-0645-2013Class II

ABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2

Horiba Instruments Incorporated / initiated 2012-09-20 / Terminated / root cause: software

Horiba Instruments, Inc dba Horiba Medical began a recall of ABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2 on . FDA classified it as Class II and gives the reason as "Horiba Medical is recalling ABX PENTRA ML because they have confirmed an issue with the ABX PENTRA ML data management and validation system. An incorrect patient result is transmitted to the Laboratory Information...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0645-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
ABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Horiba Medical is recalling ABX PENTRA ML because they have confirmed an issue with the ABX PENTRA ML data management and validation system. An incorrect patient result is transmitted to the Laboratory Information System. This result is from a previous run of the same patient that is stored in the ML database. This issue occurs in a very rare sequence of specific events. The affected software
Root cause tag
software
FDA root cause
Software design
Action
.A recall letter dated 11/09/2012 was sent to all their customers who received the ABX PENTRA ML products. The letter informed the customers of the problem identified and the action to be taken. Customers are instructed to complete and return the enclose response form within 10 days. Customers with questions about the letter are instructed to contact their local HORIBA Medical representative. Recall Letter, as produced per FCD-0377, Field Safety Notice & Acknowledgement will be mailed via FedEx to the ten (10) customers noted on the list the week of November 12, 2012. Effectivity will be evaluated by receipt of the Field Safety Notice Acknowledgment Form. Follow-up communications with customers will be conducted, as needed.
Quantity in commerce
10 units
Distribution
Nationwide in the US, including the states of CA, MI, IN, FL, and GA. Center Recall Depth is Retail.
Product codes and lots
Lot/Serial# PML0937

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0645-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0645-2013Class IITerminatedVoluntary: Firm initiated