Z-0697-2013Class II

25-Hydroxy Vitamin D EIA, Catalog #AC-57F1

IDS (Immunodiagnostic Systems Ltd.) / initiated 2012-11-07 / Terminated / root cause: manufacturing process

IDS (Immunodiagnostic Systems Ltd.) began a recall of 25-Hydroxy Vitamin D EIA, Catalog #AC-57F1. Product Usage: The IDS 25-Hydroxy Vitamin D EIA kit is an enzymeimmunoassay intended for the quantitative... on . FDA classified it as Class II and gives the reason as "Immunodiagnostic Systems is recalling the 25-Hydroxy Vitamin D EIA, AC-57F1 because the QC report was found to have an erroneous calibrator value assigned. The value of 24.4 ng/mL for calibrator 4 was stated on the QC...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0697-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
IDS (Immunodiagnostic Systems Ltd.)
Product
25-Hydroxy Vitamin D EIA, Catalog #AC-57F1. Product Usage: The IDS 25-Hydroxy Vitamin D EIA kit is an enzymeimmunoassay intended for the quantitative determination of 25-hydroxyvitamin D (25-OH D) and other hydroxylated metabolites in human serum or plasma. Results are to be used in conjunction with other clinical and laboratory data to assist the clinician in the assessment of vitamin D sufficiency in adult populations.
Product code
MRG System, Test, Vitamin D
Reason for recall
Immunodiagnostic Systems is recalling the 25-Hydroxy Vitamin D EIA, AC-57F1 because the QC report was found to have an erroneous calibrator value assigned. The value of 24.4 ng/mL for calibrator 4 was stated on the QC report, however the correct value should have read 28.4 ng/mL.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
IDS sent a recall notification letter on November 7, 2012 to all customers that received the the 25-Hydroxy Vitamin D EIA (AC-57F1). The letter informed the customers of the problem identified and the action to be taken. Customers are instructed to complete the attached Recall and Acknowledgement Form and fax back to Immunodiagnostic Systems Inc. Customers with questions are instructed to contact their local IDS office.
Quantity in commerce
675 units
Distribution
US Nationwide Distribution
Product codes and lots
Lot # 17628

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0697-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0697-2013Class IITerminatedVoluntary: Firm initiated