Z-0647-2013Class III

Immunoradiometric assay

Siemens Healthcare Diagnostics Inc. / initiated 2012-11-28 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics began a recall of Immunoradiometric assay on . FDA classified it as Class III and gives the reason as "The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0647-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Immunoradiometric assay.
Product code
JLW Radioimmunoassay, Thyroid-Stimulating Hormone
Reason for recall
The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 (level G) lot 035. The affected calibrator vials can be identified by a darker yellow color than normal. The TSI3 (level A), TSI7 (level E), and TSIX (level H)
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent Medical Device Recall letter dated November 2012 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product, complete and return the Urgent Medical Device Recall Effectiveness Check form included with the letter by fax to 302-631-7597. Customers were also instructed to determine their replacements needed, and to forward the recall letter to whomever they may have distributed the affected product. For questions customers were instructed to contact their Siemens Technical Solutions Ceter or their local technical support representative. For questions regarding this recalla call 914-524-2955.
Quantity in commerce
Lot 831=19 units (all foreign consignees) and Lot 832=2 units (all foreign consignees)
Distribution
Worldwide Distribution - USA including CA, GA, IL, IN, MI, MN, NY, and Puerto Rico. Internationally to Agentina, Bolivia, Brazil, Canada, Chile, Guatemala, South Korea, Philippines, Paraguay, Taiwan, and Uraguay,
Product codes and lots
Lots 831 and 832

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0647-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0646-2013Class IIITerminatedVoluntary: Firm initiated
Z-0647-2013Class IIITerminatedVoluntary: Firm initiated
Z-0648-2013Class IIITerminatedVoluntary: Firm initiated