Integra XKnife Stereotactic Radiosurgery and Radiotherapy Treatment Planning, Software Versions 5.0.1 and...

Integra LifeSciences Corporation / initiated 2012-12-10 / Terminated / root cause: software

Integra LifeSciences Corp. began a recall of Integra XKnife Stereotactic Radiosurgery and Radiotherapy Treatment Planning, Software Versions 5.0.1 and 5.0.2. Intended for use in stereotactic and... on . FDA classified it as Class II and gives the reason as "Depending on the system configuration, a software error message in versions 5.0.1 and 5.0.2. occurs if a beam plan is transmitted from XKnife using DICOM-RT". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1103-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra XKnife Stereotactic Radiosurgery and Radiotherapy Treatment Planning, Software Versions 5.0.1 and 5.0.2. Intended for use in stereotactic and non-stereotactic (frameless stereotactic), collimated beam, computer planned, linear accelerator (Linac) based treatment.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Depending on the system configuration, a software error message in versions 5.0.1 and 5.0.2. occurs if a beam plan is transmitted from XKnife using DICOM-RT.
Root cause tag
software
FDA root cause
Software design
Action
Integra initiated a voluntary recall on December 10, 2012 by providing a written Urgent: Medical Device Recall Notification to US consignees who have experienced the software issue due to the configuration of their system. The letter identifies the affected product, as well as, advises customers of the nature of the issue and how it affects treatment records. - Advises customers that they will be contacted when V 5.0.3 software, where the software error has been addressed, is available to schedule an upgrade. - Requests that customer consignees complete and return the Product Recall Acknowledgment Form. If customer have additional questions, they should contact the firm's service hotline at 1-888-772-7378.
Quantity in commerce
11 XKnife V5.0.1 and 1 XKnife V5.0.2
Distribution
Distributed in the states of IL, ID, LA, MA, NJ, NY, OH, PA, and SD.
Product codes and lots
Software Versions 5.0.1 and 5.0.2.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1103-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1103-2013Class IITerminatedVoluntary: Firm initiated