Space Pump Pole Clamps for the Infusomat Space Pump, model number 8713130. The Pole Clamp is an accessory to...
B. Braun Medical, Inc. began a recall of Space Pump Pole Clamps for the Infusomat Space Pump, model number 8713130. The Pole Clamp is an accessory to Space Pumps designed to support 1 to 3 pumps on a... on . FDA classified it as Class II and gives the reason as "Firm has become aware of the potential for breakage of the pole clamp locking grids, or rasters, in the Space Pole Clamps". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0623-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B Braun Medical Inc
- Product
- Space Pump Pole Clamps for the Infusomat Space Pump, model number 8713130. The Pole Clamp is an accessory to Space Pumps designed to support 1 to 3 pumps on a pole.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Firm has become aware of the potential for breakage of the pole clamp locking grids, or rasters, in the Space Pole Clamps.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- B. Braun Medical, Inc sent an Urgent Medical Device letter dated March 23, 2012 , to all affected customers. The letter identified the product the problem and the action needed by the customer to take. B. Braun has implemented an improvement to the pump locking mechanism of t he older style pole clamps and will be upgrading any older style pole clamps that may remain within your facility. Please return the completed form from attachment 1 in the self-addressed envelope provided or fax the form to (610) 849-1197 within two (2) weeks of receipt\. If you require any additional information or have any questions, comments or concerns please contact your B. Braun Medical Customer Service at 800-627-7867.
- Quantity in commerce
- 13,078
- Distribution
- Worldwide Distribution--US (nationwide) including the states of AL, AZ, NC and TN.
- Product codes and lots
- Model number 8713130 and lot numbers 209, 309, 409, 509, 809, 909, 60939747, 60943679, 60945966, 60948966, 60959578, 60960830, 60962412, 61170378, 5048020067, 5048805107, 5048920117, 5049152127, 5049484028, 5049545028, 5049614028, 5049704038, 5049788048, 5049821048, 5050123058, 5050297068, 5050385078, 5051802039, 5051879039, 5051893059, 5051940049, 5051999049, 5052519099, 5052638099, 5052690109, 5052692109, 5052711109, 5052726119, 5052788119, 5052920119, 5053045129, 5053112010, 5053129129, 5053146010, 5053207010, 5053269020, 5053449020, 5053505030, 5053534040, 5053584030, 5053605040, and 5053645040.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0623-2013 | Class II | Terminated | Voluntary: Firm initiated |