Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0695-2013Class II

Hoffman LRF Wire Bolt, Catalog number 4933-1-001 through -003

Howmedica Osteonics Corp / initiated 2012-10-25 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of Hoffman LRF Wire Bolt, Catalog number 4933-1-001 through -003 on . FDA classified it as Class II and gives the reason as "Two complaints have been filed where Hoffman LRF Wires have broken during load bearing application by patients". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0695-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Hoffman LRF Wire Bolt, Catalog number 4933-1-001 through -003.
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
Two complaints have been filed where Hoffman LRF Wires have broken during load bearing application by patients.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker sent "Urgent Medical Device Removal" notification letters and acknowledgement forms dated 10/25/2012 via Fed Ex to Branches/Agencies with return receipt. Stryker sent notification letters/acknowledgement forms dated 10/26/202 via Fed Ex to Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeons. The letters described the device issue and provided recommended actions. Questions were directed to Stryker Customer Service.
Quantity in commerce
476 units
Distribution
Nationwide Distribution, including the states of PA, OH, NY, CA and IL.
Product codes and lots
Catalog number 4933-1-001, Lot number Z02804, Z08348, Z08350, Z12549, Z16753. Catalog number 4933-1-002, Lot number Z08351, Z12554, Z16246, Z16752. Catalog number 4933-1-003 Lot Number Z08349, Z09344, Z12551, Z16996.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0695-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0695-2013Class IITerminatedVoluntary: Firm initiated