Z-0695-2013Class II
Hoffman LRF Wire Bolt, Catalog number 4933-1-001 through -003
Howmedica Osteonics Corp / initiated 2012-10-25 / Terminated / root cause: design
Stryker Howmedica Osteonics Corp. began a recall of Hoffman LRF Wire Bolt, Catalog number 4933-1-001 through -003 on . FDA classified it as Class II and gives the reason as "Two complaints have been filed where Hoffman LRF Wires have broken during load bearing application by patients". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0695-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Hoffman LRF Wire Bolt, Catalog number 4933-1-001 through -003.
- Product code
- KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
- Reason for recall
- Two complaints have been filed where Hoffman LRF Wires have broken during load bearing application by patients.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Stryker sent "Urgent Medical Device Removal" notification letters and acknowledgement forms dated 10/25/2012 via Fed Ex to Branches/Agencies with return receipt. Stryker sent notification letters/acknowledgement forms dated 10/26/202 via Fed Ex to Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeons. The letters described the device issue and provided recommended actions. Questions were directed to Stryker Customer Service.
- Quantity in commerce
- 476 units
- Distribution
- Nationwide Distribution, including the states of PA, OH, NY, CA and IL.
- Product codes and lots
- Catalog number 4933-1-001, Lot number Z02804, Z08348, Z08350, Z12549, Z16753. Catalog number 4933-1-002, Lot number Z08351, Z12554, Z16246, Z16752. Catalog number 4933-1-003 Lot Number Z08349, Z09344, Z12551, Z16996.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0695-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.