Z-0702-2013Class II

Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920

Philips North America LLC / initiated 2012-12-13 / Terminated

Philips Healthcare Inc. began a recall of Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use on . FDA classified it as Class II and gives the reason as "Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0702-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use
Product code
RCA X-Ray Controls - Fluoroscopic
Reason for recall
Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips Medical issued a notification by letter dated December 13, 2012. via Certified Mail to affected customer. The letter identified the affected product, problem and actions to be taken. Philips will contact affected customers to schedule an upgrade of the equipment . For questions or support concerning this issue, please contact Dominic Siewko at 978-659-7936.
Quantity in commerce
16 units
Distribution
US Nationwide Distribution including the states of AL, CA, CO, CT, IL, MD, MI, MO, NY, OH, PA
Product codes and lots
Serial Numbers: 10110770 11030805 11070848 11070845 11070849 11030806 11090864 11110900 12020982 114087 12030999 12061020 12041003 12061014 12061018 12081040  :

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0702-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0702-2013Class IITerminatedVoluntary: Firm initiated