Z-0702-2013Class II
Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920
Philips North America LLC / initiated 2012-12-13 / Terminated
Philips Healthcare Inc. began a recall of Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use on . FDA classified it as Class II and gives the reason as "Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0702-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use
- Product code
- RCA X-Ray Controls - Fluoroscopic
- Reason for recall
- Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Philips Medical issued a notification by letter dated December 13, 2012. via Certified Mail to affected customer. The letter identified the affected product, problem and actions to be taken. Philips will contact affected customers to schedule an upgrade of the equipment . For questions or support concerning this issue, please contact Dominic Siewko at 978-659-7936.
- Quantity in commerce
- 16 units
- Distribution
- US Nationwide Distribution including the states of AL, CA, CO, CT, IL, MD, MI, MO, NY, OH, PA
- Product codes and lots
- Serial Numbers: 10110770 11030805 11070848 11070845 11070849 11030806 11090864 11110900 12020982 114087 12030999 12061020 12041003 12061014 12061018 12081040 :