Z-0661-2013Class II

STERRAD NX Cassettes, P/N 10133. The STERRAD NX sterilizations systems are low-temperature general purpose...

Advanced Sterilization Products / initiated 2012-09-20 / Terminated / root cause: labeling and UDI

Advanced Sterilization Products began a recall of STERRAD NX Cassettes, P/N 10133. The STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture... on . FDA classified it as Class II and gives the reason as "Advanced Sterilization Products is recalling three lots of STERRAD NX System Cassettes because it does not have adequate data to support the entire duration of the product's labeled shelf-life". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0661-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Sterilization Products
Product
STERRAD NX Cassettes, P/N 10133. The STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD NX Cassettes are only used in conjunction with the STERRAD NX Sterilization system.
Product code
MLR Sterilizer, Chemical
Reason for recall
Advanced Sterilization Products is recalling three lots of STERRAD NX System Cassettes because it does not have adequate data to support the entire duration of the product's labeled shelf-life.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Advanced Sterilization Product sent an Urgent Medical Device Recall (Removal) letter dated December 4, 2012, to all affected customers. The letter informed the customers that the Sterrad NX Cassettes do not have adequate data to support the entire duration of the product's labeled shelf-life. Customers are instructed of the problems identifed and the actions to be taken. Customers are instructed to notify their customers if they have further distributed the recalled products and to have them contact Stericycle at (877) 650-7686 to arrange for return of the product. Customers with questions are instructed to call Stericycle at (877) 650-7686 Mon-Fri from 8am to 5pm, ET. For questions regarding this recall call 949-453-6400.
Quantity in commerce
12,280 units
Distribution
Worldwide Distribution - USA (nationwide) and internationally to the following countries: AR, CO, TW, SI, AU, MX, MY, HU, PH, BE, FR, DE, IE, IT, ES, NO, CH, and GB.
Product codes and lots
Lot # 11J037, 11L038, 11J033

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0661-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0661-2013Class IITerminatedVoluntary: Firm initiated