Z-0627-2013Class II

Maquet Yuno Extension Table Device Maquet GmbH & Co. KG Kehler Strabe 31 Baden-Wuerttemberg 76437...

Maquet Cardiovascular Us Sales, LLC / initiated 2012-11-02 / Terminated / root cause: design

Maquet Inc. began a recall of Maquet Yuno Extension Table Device Maquet GmbH & Co. KG Kehler Strabe 31 Baden-Wuerttemberg 76437 Rastatt,Germany Used to support the lower extremities... on . FDA classified it as Class II and gives the reason as "There is a potential issue with the screw tension of the extension device 1433.62A1 used with the YUNO table". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0627-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Maquet Yuno Extension Table Device Maquet GmbH & Co. KG Kehler Strabe 31 Baden-Wuerttemberg 76437 Rastatt,Germany Used to support the lower extremities immediately before during and after surgery. The extension device is mounted to the YUNO operating table. The product may only be operated by medically trained staff within the OR environment.
Product code
FWZ Operating Room Accessories Table Tray
Reason for recall
There is a potential issue with the screw tension of the extension device 1433.62A1 used with the YUNO table.
Root cause tag
design
FDA root cause
Device Design
Action
Maquet sent Urgent Product Field Correction letters/Field Correction Confirmation Form dated November 2, 2012 on November 15, 2012, via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that a MAQUET authorized Service Technicians would contact them to arrange an appointment to assess their device and to replace the release lever at no cost to them at their facility. Customers were asked to completed and return the enclosed confirmation form, acknowledging their receipt and understanding of the communication, in the enclosed postage prepaid envelope, by e-mailing a scanned copy to [email protected] or via fax at 973-709-6508. For questions customers were instructed to call MAQUET Technical Support at 800-777-4222, Press option2, followed by option 2. For questions regarding this recall call 973-709-7660.
Quantity in commerce
63 devices
Distribution
Worldwide Distribution - USA including AR, GA, SC, TX, TN and internationally to France, United Arab Emirates, Australia, Brazil, Canada, Switzerland, China, Germany, Spain, United Kingdom, Italy, Japan, Sweden, Thailand, Turkey, Serbia, and South Africa.
Product codes and lots
Class I exempt Device Listing Number A922956

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0627-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0627-2013Class IITerminatedVoluntary: Firm initiated