Z-0698-2013Class II

Abbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 100...

Abbott Medical / initiated 2012-12-12 / Terminated

Abbott Laboratories began a recall of Abbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 100 tests), 03L52-35 (ROW) and 03L52-37 (U.S.) (5 x 100... on . FDA classified it as Class II and gives the reason as "The Architect 25-OH Vitamin D assay has the potential to generate falsely elevated or falsely depressed results when used with certain lots of Architect Reaction Vessels". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0698-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 100 tests), 03L52-35 (ROW) and 03L52-37 (U.S.) (5 x 100 tests); in vitro diagnostic.
Product code
MRG System, Test, Vitamin D
Reason for recall
The Architect 25-OH Vitamin D assay has the potential to generate falsely elevated or falsely depressed results when used with certain lots of Architect Reaction Vessels.
FDA root cause
Mixed-up of materials/components
Action
Abbott Laboratories sent a "Product Correction Immediate Action Required" letter dated December 12, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed to immediately discontinue use and destroy any remaining inventory of the reaction vessel lots listed; to clear the ARCHITECT RV hopper of impacted RV lots following the instructions provided, prior to continued use of the ARCHITECT 25-OH Vitamin D assay; to review the patient results obtained with the affected lots with the treating physician or heath care provider to ensure they are aware of the potential analytical issue and can determine whether re-evaluation of the patient is necessary; to forward a copy of the letter to any laboratory to whom they may have forwarded the reagent; and to retain a copy of the letter for their records and to complete and return the Customer Reply Form via fax to: 1-800-777-0051 For further questions please call your local Customer Service representative at 877-422-2688.
Quantity in commerce
28,600 total
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Austria, Barbados, Belgium, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Latvia, Libya, Malaysia, Netherlands, New Zealand, Poland, Portugal, Russia, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Product codes and lots
ARCHITECT 25-OH Vitamin D Reagent Kit list 3L52, lot numbers 02412C000, exp. 12/26/2012; 02512C000, exp. 12/26/2012; 02012G000, exp. 7/16/2013; 01612G000, exp. 7/17/13; 00912G000, exp. 7/10/13; 02512I000, exp. 10/8/2013; 02612I000, exp. 10/8/2013; 03512I000, exp. 10/10/2013; 03712I000, exp. 10/10/13; when used with ARCHITECT Reaction Vessel (RV) list 7C15-01, box lot numbers (bag lot numbers): 19320P100 (6F9891501); 19344P100 (4N5891091); 19404P100 (1J1948151);  19322P100 (LM3891731); 19373P100 (6F9891521); 19407P100 (4N5948361); 19324P100 (NK8891921); 19374P100 (JG7891631); 19408P100 (5P4948471); 19340P100 (1J1890911); 19375P100 (LM3891741); 19497P100 (2L2948281); 19341P100 (2L2891001); 19379P100 (NK8891931); 20026P100 (LM3947841)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0698-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0698-2013Class IITerminatedVoluntary: Firm initiated