Z-0747-2013Class III

Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog Number: 88770 Alere Cholestech LDX...

Alere San Diego, Inc. / initiated 2012-10-22 / Terminated / root cause: design

Alere San Diego, Inc. began a recall of Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog Number: 88770 Alere Cholestech LDX Calibration Verification (hereafter referred to as CAL... on . FDA classified it as Class III and gives the reason as "Out of range (increased) HDL Cholesterol results". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0747-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alere San Diego, Inc.
Product
Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog Number: 88770 Alere Cholestech LDX Calibration Verification (hereafter referred to as CAL VERS) enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG), and glucose (GLU) test procedures on the Alere Cholestech LDX System. Calibration verification is optional for CLIA waived systems, such as the Alere Cholestech LDX System, under CLIA regulations. However local or state regulations may require that calibration verification is run at regular intervals.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Out of range (increased) HDL Cholesterol results.
Root cause tag
design
FDA root cause
Component design/selection
Action
A recall letter was sent on 10/22/12 to all customers who purchased the Alere Cholestech LDX Mutianalyte Control, 2 vial set, Catalog Number 88769, Lot #C2510 & Lot #C2510A. The letter informed the customers that these lots of controls may yield results which are out of range high for the Level 2 control when tested with the Alere Cholestech LDX HDL test. Customers are instructed of the actions to be taken. On 11/16/12, the recall expanded to include the Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog 88770, Lot#271097. The problem identified is that these lots of calibration verification materials may yield results which are out of range high when tested with the Alere Cholestech LDX HDL test. Customers are instruted of the actions to be taken. On 3/6/13, the recall expanded to include the Alere Cholestech LDX Calibration Verification, 4 Vial set, Catalog # 88770, Lot#298148 and 301886. The recall expanded to include the Alere Cholestech LDX Multianalyte Control, 2 vial Set, Catalog # 88769, Lots#C2572, C2572A, C2593 and C2593A. Customers are instructed to complete and fax the enclosed verification form within 10 days. Customers with questions are instructed to contact Alere San Diego, 9975 Summers Ridge Rd, San Diego, CA 92121, (877) 308-8289, fax (858) 805-8457, and [email protected].
Quantity in commerce
614 kits
Distribution
Worldwide Distribution - USA (nationwide) Puerto Rico and South Korea.
Product codes and lots
Lot#271097, 298148, and 301886

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0747-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0746-2013Class IIITerminatedVoluntary: Firm initiated
Z-0747-2013Class IIITerminatedVoluntary: Firm initiated