Z-0687-2013Class II

NEOBASE NON-DERIVATIZED MSMS KIT Product Number: 3040-001U. This tandem mass spectrometry (MSMS) reagent kit...

Revvity Health Sciences, Inc. / initiated 2012-11-26 / Terminated / root cause: manufacturing process

Perkinelmer began a recall of NEOBASE NON-DERIVATIZED MSMS KIT Product Number: 3040-001U. This tandem mass spectrometry (MSMS) reagent kit is intended for the measurement and evaluation of... on . FDA classified it as Class II and gives the reason as "Truncated V-Bottomed Microplate Wells used in assays contain defective wells". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0687-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Revvity Health Sciences, Inc.
Product
NEOBASE NON-DERIVATIZED MSMS KIT Product Number: 3040-001U. This tandem mass spectrometry (MSMS) reagent kit is intended for the measurement and evaluation of amino acids, succinylacetone, free carnitine, and acylcarnitine concentrations from newborn heel prick blood samples dried on filter paper. Quantitative analysis of these analytes ) and their relationship with each other is intended to provide analyte concentration profiles that may aid in screening newborns for metabolic disorders.
Product code
NQL System, Test, Amino Acids, Free Carnitines And Acylcarnitines Tandem Mass Spectrometry
Reason for recall
Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Perkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually detected in a well of the plate, it is required during use of the well, the volume of solution in the well should be visually confirmed to be consistent before and after incubation. Please contact PerkinEimer Technical Support for further information (800) 321 -9632, select prompt 2.
Quantity in commerce
819 kits (10 plates/kit)
Distribution
Nationwide Distribution.
Product codes and lots
Lot Numbers: 617253 , 617346, 618494 , 618653, 618764 , 619453

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0687-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0686-2013Class IITerminatedVoluntary: Firm initiated
Z-0687-2013Class IITerminatedVoluntary: Firm initiated
Z-0688-2013Class IITerminatedVoluntary: Firm initiated
Z-0689-2013Class IITerminatedVoluntary: Firm initiated
Z-0690-2013Class IITerminatedVoluntary: Firm initiated