Z-0688-2013Class II
NEONATAL TOTAL GALACTOSE KIT Product Numbers: 3029-0010, 3029-001B. This kit is intended for the quantitative...
Revvity Health Sciences, Inc. / initiated 2012-11-26 / Terminated / root cause: manufacturing process
Perkinelmer began a recall of NEONATAL TOTAL GALACTOSE KIT Product Numbers: 3029-0010, 3029-001B. This kit is intended for the quantitative determination of total galactose (galactose and... on . FDA classified it as Class II and gives the reason as "Truncated V-Bottomed Microplate Wells used in assays contain defective wells". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0688-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Revvity Health Sciences, Inc.
- Product
- NEONATAL TOTAL GALACTOSE KIT Product Numbers: 3029-0010, 3029-001B. This kit is intended for the quantitative determination of total galactose (galactose and galactose-1-phosphate) concentrations in blood specimens dried on filter paper as an aid in screening newborns for galactosemia.
- Product code
- JLA Conversion To Creatinine, Creatine
- Reason for recall
- Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Perkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually detected in a well of the plate, it is required during use of the well, the volume of solution in the well should be visually confirmed to be consistent before and after incubation. Please contact PerkinEimer Technical Support for further information (800) 321 -9632, select prompt 2.
- Quantity in commerce
- 3029-0010 =1562 kits (10 plats/kit) and 3029-001B=13 kits (50plates/kit)
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lot Numbers: 618201, 619200, 618263
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0686-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0687-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0688-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0689-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0690-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.