Z-0690-2013Class II
Microplate, truncated V-bottomed lots included in the above listed products Product Number: 1380 5465. DBS...
Revvity Health Sciences, Inc. / initiated 2012-11-26 / Terminated / root cause: manufacturing process
Perkinelmer began a recall of Microplate, truncated V-bottomed lots included in the above listed products Product Number: 1380 5465. DBS Microplate is a general purpose 96 well microplate... on . FDA classified it as Class II and gives the reason as "Truncated V-Bottomed Microplate Wells used in assays contain defective wells". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0690-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Revvity Health Sciences, Inc.
- Product
- Microplate, truncated V-bottomed lots included in the above listed products Product Number: 1380 5465. DBS Microplate is a general purpose 96 well microplate used in the preparation and examination of specimens from the human body.
- Reason for recall
- Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Perkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually detected in a well of the plate, it is required during use of the well, the volume of solution in the well should be visually confirmed to be consistent before and after incubation. Please contact PerkinEimer Technical Support for further information (800) 321 -9632, select prompt 2.
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Lot Numbers: 616792 , 617141 , 617980 , 618319, 618664, 618831, 619175
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0686-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0687-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0688-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0689-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0690-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.