Ortho Clinical Diagnostics VITROS Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS...
Ortho-Clinical Diagnostics began a recall of Ortho Clinical Diagnostics VITROS Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS Vapor Adsorption Cartridges are used to remove... on . FDA classified it as Class II and gives the reason as "Some VITROS Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0755-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- Ortho Clinical Diagnostics VITROS Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS Vapor Adsorption Cartridges are used to remove vapor contaminants that may potentially interfere with Signal Reagent used with MicroWell Assays on the VITROS 3600, VITROS 5600, and VITROS ECi/ECiQ Systems.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Some VITROS Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently installed on the VITROS System instead of the intended VITROS Vapor Adsorption Cartridge which may bias results.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Ortho-Clinical Diagnostics sent URGENT PRODUCT CORRECTION Letters (dated 12/13/12) along with Q&A and Confirmation of Receipt Forms to distributors and consignees on 12/13/12. They were advised of the issue and instructed to discard their current inventory of VITROS Vapor Adsorption Cartridges. On 12/13/12, foreign affiliates were informed by e-mail of the issue and instructed to notify their consignees of the issue and actions.
- Quantity in commerce
- Domestic: 15,891; Foreign: 5237
- Distribution
- Worldwide Distribution, including Nationwide (USA) and the countries of AU, GM, FR, IT, SP, BR, BM, AR, CA, JA, MX, UK, CH, CL, CO, IN, EC, PE, PY, PA, PT, NZ, SG, VE and UY.
- Product codes and lots
- CAT No. 680 0100; the cartridges do not have lot numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0755-2013 | Class II | Terminated | Voluntary: Firm initiated |