Endotrig Trigger Release Sterile Micro Hook Blade
Instratek, Incorporated began a recall of Endotrig Trigger Release Sterile Micro Hook Blade Product Usage: The blade is intended to release A-1 pulley in endoscopic trigger release procedures on . FDA classified it as Class II and gives the reason as "Instratek, Inc.has received multiple complaints of the 1052 Trigger Finger Release Blades which could not be advanced into the cannula through which it is used during trigger finger release procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0922-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Instratek, Incorporated
- Product
- Endotrig Trigger Release Sterile Micro Hook Blade Product Usage: The blade is intended to release A-1 pulley in endoscopic trigger release procedures.
- Product code
- NBH Accessories, Arthroscopic
- Reason for recall
- Instratek, Inc.has received multiple complaints of the 1052 Trigger Finger Release Blades which could not be advanced into the cannula through which it is used during trigger finger release procedures.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm decided to recall the product and sent notification letters to consignees dated 09/18/2012.
- Quantity in commerce
- 235
- Distribution
- USA Nationwide Distribution including the states of: AZ, MI, NE, NV, and TX.
- Product codes and lots
- Model number 1052; Lots M674180, MAKL170, M674200
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0922-2013 | Class II | Terminated | Voluntary: Firm initiated |