Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0922-2013Class II

Endotrig Trigger Release Sterile Micro Hook Blade

Instratek, Incorporated / initiated 2012-09-18 / Terminated / root cause: manufacturing process

Instratek, Incorporated began a recall of Endotrig Trigger Release Sterile Micro Hook Blade Product Usage: The blade is intended to release A-1 pulley in endoscopic trigger release procedures on . FDA classified it as Class II and gives the reason as "Instratek, Inc.has received multiple complaints of the 1052 Trigger Finger Release Blades which could not be advanced into the cannula through which it is used during trigger finger release procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0922-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Instratek, Incorporated
Product
Endotrig Trigger Release Sterile Micro Hook Blade Product Usage: The blade is intended to release A-1 pulley in endoscopic trigger release procedures.
Product code
NBH Accessories, Arthroscopic
Reason for recall
Instratek, Inc.has received multiple complaints of the 1052 Trigger Finger Release Blades which could not be advanced into the cannula through which it is used during trigger finger release procedures.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm decided to recall the product and sent notification letters to consignees dated 09/18/2012.
Quantity in commerce
235
Distribution
USA Nationwide Distribution including the states of: AZ, MI, NE, NV, and TX.
Product codes and lots
Model number 1052; Lots M674180, MAKL170, M674200

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0922-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0922-2013Class IITerminatedVoluntary: Firm initiated