Z-0699-2013Class II

Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits...

Becton Dickinson & Co / initiated 2012-12-13 / Terminated

Becton Dickinson & Co. began a recall of Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological... on . FDA classified it as Class II and gives the reason as "In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0699-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological testing. The BBL Crystal Enteric/Nonfermentor (E/NF) identification (ID) System is for the identification of aerobic gram-negative bacteria that belong to the family Enterobacteriaceae as well as some of the more frequently isolated glucose fermenting and nonfermenting gram-negative bacilli.
Product code
JSS Kit, Identification, Enterobacteriaceae
Reason for recall
In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification.
FDA root cause
Mixed-up of materials/components
Action
Becton Dickinson initiated the recall by letter to direct accounts on 12/13/2012. Ex-US BD sites were notified via email, distributors were notified by fax and customer letters were delivered by UPS ground. The notification advised that the recalled kit contained citrate (CIT} and malonic acid (MLO} reagents in the reversed prong/well positions. Customers were advised to discontinue use of identified lot numbers and discard any remaining packages for replacements, and also to return the enclosed reply form. The recalling firm requested distributors to submit their customer list for direct BD notification to all end users. Contact the BD Customer Service Department at 1-800-638-8663 if further assistance is required.
Quantity in commerce
67,820 units
Distribution
Worldwide Distribution-USA including the states of MA, TN, KY, IN, GA, NC, TX, NY, FL, AZ, WA, NJ, MI, and CA, and the countries of Uruguay, Taiwan, Singapore, Australia, South Korea, Mexico, Hong Kong, Guatemala, Colombia, Chile, Canada, India, Japan, Brazil, China and Belgium.
Product codes and lots
Lot Number/ Exp Date 2086315 February 20, 2013 2101435 February 20, 2013 2104316 February 27, 2013 2129460 February 27, 2013 2129487 March 13, 2013 2163211 February 27, 2013 2163213 March 20, 2013 2163215 March 27, 2013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0699-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0699-2013Class IITerminatedVoluntary: Firm initiated