Z-0810-2013Class II

Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar

Mizuho Orthopedic Systems Inc / initiated 2012-12-17 / Terminated / root cause: manufacturing process

Mizuho OSI began a recall of Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar on . FDA classified it as Class II and gives the reason as "Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0810-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mizuho Orthopedic Systems Inc
Product
Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.
Product code
JEA Table, Surgical With Orthopedic Accessories, Ac-Powered
Reason for recall
Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Mizuho sent a Recall Notification letter dated December 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The Orthopedic Table Spar will be replaced at no cost by Mizuho technicians. Customers were informed the potential failure and were advised inspection of tables prior to use. If tables are experiencing the described failure, contact Mizuho Resource Group at 1-800-777-4674, option 2. .
Quantity in commerce
869 devices
Distribution
USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.
Product codes and lots
Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0810-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0810-2013Class IITerminatedVoluntary: Firm initiated