Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar
Mizuho OSI began a recall of Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar on . FDA classified it as Class II and gives the reason as "Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0810-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mizuho Orthopedic Systems Inc
- Product
- Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.
- Product code
- JEA Table, Surgical With Orthopedic Accessories, Ac-Powered
- Reason for recall
- Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Mizuho sent a Recall Notification letter dated December 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The Orthopedic Table Spar will be replaced at no cost by Mizuho technicians. Customers were informed the potential failure and were advised inspection of tables prior to use. If tables are experiencing the described failure, contact Mizuho Resource Group at 1-800-777-4674, option 2. .
- Quantity in commerce
- 869 devices
- Distribution
- USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.
- Product codes and lots
- Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0810-2013 | Class II | Terminated | Voluntary: Firm initiated |