Z-0913-2013Class II

Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes...

Siemens Healthcare Diagnostics Inc. / initiated 2012-11-28 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes such as Glucose, Magnesium, and Potassium on . FDA classified it as Class II and gives the reason as "Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0913-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter AU5800 Clinical Chemistry Analyzer. An automated chemistry analyzer that measures analytes such as Glucose, Magnesium, and Potassium.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Due to a software data storage limitation which only occurs under certain conditions, measurements of the Reagent Blank (RB) or Calibration (CAL) sample results are not saved and the software reverts to a previously stored, acceptable RB/CAL result.
Root cause tag
software
FDA root cause
Software design
Action
Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated November 28, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the Call Center Hotline at (800) 854-3633 for questions regarding the notice.
Quantity in commerce
28 in U.S.; 4 Canada; 318 to other foreign countries.
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Argentina, Australia, Austria, Brazil, China, Czech Republic, France, Germany, Hungary, India, Ireland, Israel, Italy, Korea Republic, Malaysia, Netherlands, Russian Federation, Spain, Switzerland, Taiwan, Turkey, United Kingdom, and Viet Nam.
Product codes and lots
All serial numbers and all software versions 4.05 and prior.  Part# Item# A94907 AU5811-02 A94912 AU5821-02 A94917 AU5831-02 A94922 AU5841-02 A94908 AU5811-03 A94913 AU5821-03 A94918 AU5831-03 A94923 AU5841-03 A94909 AU5811-04 A94914 AU5821-04 A94919 AU5831-04 A94924 AU5841-04 A94910 AU5811-06 A94915 AU5821-06 A94920 AU5831-06 A94925 AU5841-06

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0913-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0913-2013Class IITerminatedVoluntary: Firm initiated