Z-0787-2013Class II

Exactech NOVATION CROWN CUP, NEUTRAL LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 130-32-53...

Exactech, Inc. / initiated 2012-11-05 / Terminated / root cause: labeling and UDI

Exactech, Inc. began a recall of Exactech NOVATION CROWN CUP, NEUTRAL LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 130-32-53. Exactech Hip Systems are indicated for use in... on . FDA classified it as Class II and gives the reason as "Recall of Novation Crown Cup Connection GXL Neutral Liners Catalog No. (130-32-53) and Novation Crown Cup Connection GXL + 5mm Lateralized Liners Catalog No. (136-32-53). Exactech discovered the device labeling was...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0787-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Exactech, Inc.
Product
Exactech NOVATION CROWN CUP, NEUTRAL LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 130-32-53. Exactech Hip Systems are indicated for use in surgery for hip replacement.
Product code
LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Reason for recall
Recall of Novation Crown Cup Connection GXL Neutral Liners Catalog No. (130-32-53) and Novation Crown Cup Connection GXL + 5mm Lateralized Liners Catalog No. (136-32-53). Exactech discovered the device labeling was inadvertently transposed.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Exactech, Inc. notified consignees by telephone on 11/05/12 of the affected devices and sent an "IMPORTANT PRODUCT RECALL NOTICE" letter dated 11/06/12. The letter describes the purpose of the recall, the product affected, the problem and actions to be taken by the consignees. A Product Recall Response Form was attached for the consignees to contact the Exactech inventory representative within 5 business days to confirm quantities of product in their inventory and return them to Exactech. Consignees with any questions relating to these devices should call 1-800-392-2832.
Quantity in commerce
26 Units
Distribution
Distributed in the states of TN, FL, VA, WA and NV.
Product codes and lots
Catalog No. 130-3253, Serial Numbers 2479944 - 2479969 and Catalog No. 136-32-53, Serial Numbers 2479873 - 2479898.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0787-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0787-2013Class IITerminatedVoluntary: Firm initiated
Z-0788-2013Class IITerminatedVoluntary: Firm initiated