Z-0785-2013Class II
Solara 3G Custom Manual Wheelchair
Invacare GmbH / initiated 2012-09-10 / Terminated / root cause: manufacturing process
Invacare Corporation began a recall of Solara 3G Custom Manual Wheelchair Product Usage: To provide mobility to a person restricted to a sitting position on . FDA classified it as Class II and gives the reason as "The potential exists for the wheel to rotate freely despite engagement of hub brake". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0785-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Invacare GmbH
- Product
- Solara 3G Custom Manual Wheelchair Product Usage: To provide mobility to a person restricted to a sitting position.
- Product code
- IOR Wheelchair, Mechanical
- Reason for recall
- The potential exists for the wheel to rotate freely despite engagement of hub brake.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Invacare sent an URGENT: RECALL INFORMATION Letter dated September 10, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to 1) Complete the Provider Response Card indicating your receipt and understanding of these instructions. 2) Contact your customers to inform them of the recall. It is recommended that you provide them with the enclosed Customer Letter. Set up an appointment to correct the customer's wheelchair as soon as possible. 3) Please examine the wheelchair to confirm it has one of the serial numbers listed above. The serial number can be found on the front lower crossbar located on the right while sitting in the chair. 4) Using the enclosed retrofit kit, make the necessary correction. Installation instructions can be found on page 109 of the Solara service manual, which is enclosed. 5) Document the correction of the wheelchair on the enclosed tracking sheet. 6) Using the enclosed pre-paid UPS mailing label and original box, return the replaced parts to the address listed on the label. 7) Fax or email the Provider Response Form and tracking sheet to 330-329-3458 or [email protected]. For questions please contact Invacare Customer Service at 800-333-6900.
- Quantity in commerce
- 3 units
- Distribution
- US Nationwide Distribution including the states of CA & MT.
- Product codes and lots
- Model #3G, Serial numbers: 12GE003750, 12GE003554 and 12HE001766.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0785-2013 | Class II | Terminated | Voluntary: Firm initiated |