Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0667-2013Class II

Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The...

Terumo Cardiovascular Systems Corporation / initiated 2013-01-07 / Terminated / root cause: manufacturing process

Terumo Cardiovascular Systems Corporation began a recall of Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns" High Flow Aortic Arch Cannula... on . FDA classified it as Class II and gives the reason as "Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0667-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
TERUMO sent an URGENT MEDICAL DEVICE RECALL notification to all consignees on January 7, 2013. Terumo CVS is advising users to discontinue use and return all affected cannulae in inventory. Terumo CVS will issue credit for returned cannulae. Terumo CVS providied a list of marketed devices that can be used as alternatives. Contact Terumo CVS Customer Service with questions or concerns: 800- 521- 2818 Monday Friday, 8 AM - 6 PM EST For questions regarding this recall call 734-741-6173.
Quantity in commerce
6830 total all products
Distribution
Nationwide and worldwide: AL , AR , AZ , CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , ME , MI, MO , NC , ND , NV , NY , OH , OK , PA , SD , TN , TX , UT , VA , WA , and WI. Bahrain, BRAZIL, CANADA, CHILE, Honduras, Singapore, Taiwan, Trinidad and Tobago, and UNITED ARAB EMIRATES (UAE).
Product codes and lots
Lot number: 0677284 and Catalog number: 13020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0667-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0666-2013Class IITerminatedVoluntary: Firm initiated
Z-0667-2013Class IITerminatedVoluntary: Firm initiated
Z-0668-2013Class IITerminatedVoluntary: Firm initiated
Z-0669-2013Class IITerminatedVoluntary: Firm initiated
Z-0670-2013Class IITerminatedVoluntary: Firm initiated
Z-0671-2013Class IITerminatedVoluntary: Firm initiated