Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The...
Terumo Cardiovascular Systems Corporation began a recall of Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns" High Flow Aortic Arch Cannula... on . FDA classified it as Class II and gives the reason as "Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0667-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Sarns flexible arterial cannula, 7.3 mm (22 Fr) OD with 3/8in connector, suture ring, 9.5in (24 cm) long The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery
- Product code
- DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
- Reason for recall
- Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- TERUMO sent an URGENT MEDICAL DEVICE RECALL notification to all consignees on January 7, 2013. Terumo CVS is advising users to discontinue use and return all affected cannulae in inventory. Terumo CVS will issue credit for returned cannulae. Terumo CVS providied a list of marketed devices that can be used as alternatives. Contact Terumo CVS Customer Service with questions or concerns: 800- 521- 2818 Monday Friday, 8 AM - 6 PM EST For questions regarding this recall call 734-741-6173.
- Quantity in commerce
- 6830 total all products
- Distribution
- Nationwide and worldwide: AL , AR , AZ , CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , ME , MI, MO , NC , ND , NV , NY , OH , OK , PA , SD , TN , TX , UT , VA , WA , and WI. Bahrain, BRAZIL, CANADA, CHILE, Honduras, Singapore, Taiwan, Trinidad and Tobago, and UNITED ARAB EMIRATES (UAE).
- Product codes and lots
- Lot number: 0677284 and Catalog number: 13020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0666-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0667-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0668-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0669-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0670-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0671-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY