Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to...
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity... on . FDA classified it as Class III and gives the reason as "The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0933-2013
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Total Iron Binding Capacity (IBCT) Flex Reagent Cartridge. An in vitro diagnostic test intended to quantitatively measure Total Iron Binding Capacity in human serum.
- Product code
- JMO Ferrozine (Colorimetric) Iron Binding Capacity
- Reason for recall
- The analytical sensitivity claim of 6 ug/dL (1.1 umol/L) in the Instructions for Use (IFU) is no longer valid. Siemens is conducting a voluntary corrective action for Dimension Total Iron Binding Capacity (IBCT) to modify the analytical sensitivity claim to 36 ug/dL (6.44 umol/L).
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Siemens Healthcare sent a "Urgent Medical Device Correction" letter dated Dec 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were notified that theanalytical sensitivity of 6 ug/dL (1.1 umol/L) in the Dimension(R) IBCT FLex(R) Reagent Cartridge instructions for us (IFU) is no longer valid. Please make adjustments to middle wave products or Laboratory Information Systems ( LIS ) as appropriate for your Laboratory. Please complete the attached form and fax it to (302) 631-8467, to indicate you have received this information. If you have any questions please contact your Siemens Technical Solutions Center or your local Siemens technicial support representative at (302) 631-6311.
- Quantity in commerce
- 19, 503 cartons
- Distribution
- Nationwide Distribution
- Product codes and lots
- Siemens Material Number (SMN) 10444944, Catalog number DF84 -- lot numbers BA3016, EXP 2013-01-16; EC3045, EXP 2013-02-14; EA3073, EXP 2013-03-14; DC3102, EXP 2013-04-12; FC3122, EXP 2013-05-02; EA3157, EXP 2013-06-06; BB3184, EXP 2013-07-03; EC3219, EXP 2013-08-07; EC3220, EXP 2013-08-08; GA3248, 2013-09-05; and FB3290, EXP 2013-10-17
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0933-2013 | Class III | Terminated | Voluntary: Firm initiated |