Z-0788-2013Class II

Exactech NOVATION CROWN CUP +5mm LATERALIZED LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF...

Exactech, Inc. / initiated 2012-11-05 / Terminated / root cause: labeling and UDI

Exactech, Inc. began a recall of Exactech NOVATION CROWN CUP +5mm LATERALIZED LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 136-32-53. Exactech Hip Systems are indicated for use... on . FDA classified it as Class II and gives the reason as "Recall of Novation Crown Cup Connection GXL Neutral Liners Catalog No. (130-32-53) and Novation Crown Cup Connection GXL + 5mm Lateralized Liners Catalog No. (136-32-53). Exactech discovered the device labeling was...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0788-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Exactech, Inc.
Product
Exactech NOVATION CROWN CUP +5mm LATERALIZED LINER, Use with Crown Cup Only, GXL UHMWPE, SN 0123456, REF 136-32-53. Exactech Hip Systems are indicated for use in surgery for hip replacement.
Product code
LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Reason for recall
Recall of Novation Crown Cup Connection GXL Neutral Liners Catalog No. (130-32-53) and Novation Crown Cup Connection GXL + 5mm Lateralized Liners Catalog No. (136-32-53). Exactech discovered the device labeling was inadvertently transposed.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Exactech, Inc. notified consignees by telephone on 11/05/12 of the affected devices and sent an "IMPORTANT PRODUCT RECALL NOTICE" letter dated 11/06/12. The letter describes the purpose of the recall, the product affected, the problem and actions to be taken by the consignees. A Product Recall Response Form was attached for the consignees to contact the Exactech inventory representative within 5 business days to confirm quantities of product in their inventory and return them to Exactech. Consignees with any questions relating to these devices should call 1-800-392-2832.
Quantity in commerce
26 Units
Distribution
Distributed in the states of TN, FL, VA, WA and NV.
Product codes and lots
Catalog No. 130-3253, Serial Numbers 2479944 - 2479969 and Catalog No. 136-32-53, Serial Numbers 2479873 - 2479898.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0788-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0787-2013Class IITerminatedVoluntary: Firm initiated
Z-0788-2013Class IITerminatedVoluntary: Firm initiated