Z-0962-2013Class II
DePuy CONFIDENCE SPINAL CEMENT SYSTEM 11C Plus KIT Catalog Number: 283910000. Delivery of CONFIDENCE Spinal...
DePuy Spine, Inc. / initiated 2013-01-15 / Terminated / root cause: manufacturing process
DePuy Spine, Inc. began a recall of DePuy CONFIDENCE SPINAL CEMENT SYSTEM 11C Plus KIT Catalog Number: 283910000. Delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of... on . FDA classified it as Class II and gives the reason as "Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0962-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DePuy Spine, Inc.
- Product
- DePuy CONFIDENCE SPINAL CEMENT SYSTEM 11C Plus KIT Catalog Number: 283910000. Delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures.
- Product code
- NDN Cement, Bone, Vertebroplasty
- Reason for recall
- Cement mixer used to prepare cement for surgery may not turn properly and may jam or stuck
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- DePuy Synthes Spine issued notification letter dated 1/11/13 issued via UPS Next Day Air Saver on 17 January 2013. Accounts requested to review inventory, quarantine any products listed in this recall notice and to complete and return the enclosed Business Reply Form. Contact the firm at 508-977-6606 for questions relating to this notice.
- Quantity in commerce
- 7773
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Argentina, Armenia, Kazakhstan, Russia, Ukraine, Australia, Austria Belgium, Luxembourg , Brazil, Canada, China, Colombia, Cyprus, Jordan Lebanon, Saudi Arabia, UAE, Czech Republic, Finland ,France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Latvia, Malaysia Mexico, Netherlands, New Zealand, Norway, Sweden, Poland, Portugal Singapore, Slovakia, Slovenia, South Africa, Spain, Switzerland, Turkey, and UK.
- Product codes and lots
- All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0960-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0961-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0962-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0963-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.