Z-0767-2013Class II

Phillips Easy Diagnost Eleva DRF. Model Number: 706034

Philips North America LLC / initiated 2013-01-16 / Terminated / root cause: software

Philips Healthcare Inc. began a recall of Phillips Easy Diagnost Eleva DRF. Model Number: 706034 Product Usage: Universal Fluoroscopic Applications on . FDA classified it as Class II and gives the reason as "The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0767-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Phillips Easy Diagnost Eleva DRF. Model Number: 706034 Product Usage: Universal Fluoroscopic Applications
Product code
KPR System, X-Ray, Stationary
Reason for recall
The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
Phillips issued an Urgent Field Safety Notice to all customers on January 14, 2013, to notify users that the Phillips EasyDiagnost Eleva systems, release 3.0 has a mirror icon that could be misinterpereted and could potentially lead to misdiagnosis. Philips will upgrade the system to version 3.1 which involves software and hardware upgrades of the system. This software provides a replacement icon that reduces the risk of the potential misinterpretation of the original mirror icon. Additionally, Philips will provide updated Instructions for Use (IFU) for the Eleva workspot and the EasyDiagnost 3.1 system and an addendum to the EasyDiagnost system IFU. Users will be contacted by Phillips to schedule the installation. For questions contact your local Philips representative: 1-800-722-9377.
Quantity in commerce
14 units
Distribution
US Nationwide Distribution including the states of FL, CA, WA, AZ, NC, KY, DE, AZ and MO.
Product codes and lots
Serial Numbers 362600, 362601, 364209, 364210, 375245, 375850, 376439, 384396, 387492, 398185, 399440, 399621, 399623, 402409

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0767-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0767-2013Class IITerminatedVoluntary: Firm initiated