Phillips Easy Diagnost Eleva DRF. Model Number: 706034
Philips Healthcare Inc. began a recall of Phillips Easy Diagnost Eleva DRF. Model Number: 706034 Product Usage: Universal Fluoroscopic Applications on . FDA classified it as Class II and gives the reason as "The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0767-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Phillips Easy Diagnost Eleva DRF. Model Number: 706034 Product Usage: Universal Fluoroscopic Applications
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- Phillips issued an Urgent Field Safety Notice to all customers on January 14, 2013, to notify users that the Phillips EasyDiagnost Eleva systems, release 3.0 has a mirror icon that could be misinterpereted and could potentially lead to misdiagnosis. Philips will upgrade the system to version 3.1 which involves software and hardware upgrades of the system. This software provides a replacement icon that reduces the risk of the potential misinterpretation of the original mirror icon. Additionally, Philips will provide updated Instructions for Use (IFU) for the Eleva workspot and the EasyDiagnost 3.1 system and an addendum to the EasyDiagnost system IFU. Users will be contacted by Phillips to schedule the installation. For questions contact your local Philips representative: 1-800-722-9377.
- Quantity in commerce
- 14 units
- Distribution
- US Nationwide Distribution including the states of FL, CA, WA, AZ, NC, KY, DE, AZ and MO.
- Product codes and lots
- Serial Numbers 362600, 362601, 364209, 364210, 375245, 375850, 376439, 384396, 387492, 398185, 399440, 399621, 399623, 402409
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0767-2013 | Class II | Terminated | Voluntary: Firm initiated |