GE Healthcare Definium 5000/5220493-2 X-Ray System
GE Healthcare, LLC began a recall of GE Healthcare Definium 5000/5220493-2 X-Ray System on . FDA classified it as Class II and gives the reason as "When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0704-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Definium 5000/5220493-2 X-Ray System
- Product code
- JAC System, X-Ray, Film Marking, Radiographic
- Reason for recall
- When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The exposure can still be made as commanded, however it may not be properly indicated.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE Healthcare will remedy the issue or bring the product into compliance with each applicable Federal standard. A GE Healthcare Service Representative will update the software on the system to address the issue. Call Center phone number: 800-437-1171 (United States). For other countries, please contact your local GE Healthcare Service Representative if you have any questions or concerns regarding this notification. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately per the contact information above.
- Quantity in commerce
- 312 installed in US
- Distribution
- Nationwide Distribution
- Product codes and lots
- Definium Model 5000/5220493-2
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0703-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0704-2013 | Class II | Terminated | Voluntary: Firm initiated |