Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with...
Ad-Tech Medical Instrument Corporation began a recall of Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation equipment, for... on . FDA classified it as Class I and gives the reason as "The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0892-2013
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AD-Tech Medical Instrument Corp.
- Product
- Ad-Tech Medical Instrument Corporation Macro Micro Subdural Electrode. For temporary (<30 days) use with recording, monitoring and stimulation equipment, for the recording, monitoring and stimulation of electrical signals on the surface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.
- Product code
- GYC Electrode, Cortical
- Reason for recall
- The microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is explanted. Recall expanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- AD-TECH Medical Instrument Corporation sent an "Urgent Medical Device Recall" letter dated December 18, 2012 to all affected consignees. The letter described the Reason for the Voluntary Recall, Risk to Health, Actions to be taken by the Customer, Product and Distribution Information, Type of Action by Ad-Tech Medical Instrument Corporation and Other Information. Consignees were instructed to quarantine the affected product and contact the firm for a Return Material Authorization number. If the product have been further distributed consignees were advised to notify their customers of the recall. The letter requested consignees to return the Acknowledgement and Receipt Form. For questions contact your Ad-Tech Clinical Specialist. Expanded recall letters dated May 2, 2013 were sent out April 29, 2013. The letters were formatted as the initial ones sent on 12/18/2013.
- Quantity in commerce
- 144 (103 US; 41 OUS)
- Distribution
- Worldwide Distribution - USA Nationwide including the states of CA, GA, LA, MA, MD, MI, MN, NH, PA, RI, TX, UT, WI, NY IA, WI and the countries of: Australia, CANADA, ENGLAND, JAPAN, KOREA, SPAIN, FRANCE, and SWITZERLAND.
- Product codes and lots
- REF numbers (catalog numbers): EG10A-SP10N-000, EG34B-SP10X-000, EGD7A-MP12N-000, EGE4A-SP10N-000, EGE4A-SP10X-000, QG10B-SP10X-000, QG10C-SP10N-000, QG10D-SP10N-000, QG11A-SP10X-000, QG16A-SP10X-000, QG17A-SP10X-000, QG18A-SP10N-000, QG18A-SP10X-000, QG20B-SP10N-000, QG20B-SP10X-000, QG20C-SP10N-000, QG20C-SP10X-000, QG20D-SP10X-200, QG20E-SP10X-000, QG20F-SP10X-000, QG24A-SP10N-000, QG26A-SP10N-000, QG32A-SP10X-000, QG32B-SP10X-000, QG36A-SP10N-000, QG36A-SP10X-000, QG36B-SP10N-000, QG44A-SP10X-000, QG53A-SP10X-000, QG66A-SP85N-000, QG69A-SP10N-000, QG88A-SP10X-000, QG96A-SP10N-000, QG96A-SP10X-000, QG96C-SP10X-000, QG96D-SP10X-000, QGC6A-SP07N-000, QGC8A-SP10N-000, QGC8A-SP10X-000, QGC8B-SP10X-000, QGC8C-SP10X-000, QGC8D-SP10N-000, QGC8E-SP10X-000, QGC8F-SP10X-000, QS20A-SP10N-000, QS40B-SP10X-000, Lot numbers: 208140314, 208140343, 208140344, 208140353, 208140354, 208140362, 208140369, 208140374, 208140381, 208140382, 208140387, 208140390, 208140395, 208140398, 208140399, 208140414, 208140417, 208140421, 208140431, 208140443, 208140445, 208140450, 208140452, 208140453, 208140458, 019020, 208140356R, 208140357R, 208140452R, C010180001, C702201001, C703160006, C704040001, C800250002, C800350004, C801240003, C802220003, C802220005, C802220006, C805210003, C805210004, C908221002, G012111001, G012140002, G015020002, G016240001, G017250004, G017250005, G017250006, G018011001, G018011002, G018011003, G019021002, G019021003, G019150003, G112270004, G114160001, G114160002, G115080002, G115170004, G115280005, G119020004, G211220001, G608060002, G704150001, G805270001, G900211002, G903011001, G903080003, S903080004, S905280003. Recall exanded on 5/2/2013 to include additional Macro Micro Subdural Electrodes: REF numbers (catalog numbers): CG-64BPX-F436, CMMS-22PX-F478 CG-32BPX-F412, CGIE-98BPX-F341, CGIE-82BPN-F377, CMMS-69PN-F480, CGIE-98BPN-F341, CG-9SN-F496, CG-32BPX-F426, CG-16BPX-F445, CG-16BPX-F446, CG-256PX-F425, CGIE-39BPX-F442, CG-46BPX-F465, CG-64BPX-F419, CG-32BPX-F418. Lot numbers: 20814044, 208140447, 208140463, 208140463, 208140476, 208140463, 08140463, 208140424, 208140439, G119221001,G212040002, G216040002, 218190, 208140431, 208140450, 208140432, 208140433, 208140420, 208140437, 208140450, 208140422, 208140422, 208140422, 208140422
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0892-2013 | Class I | Terminated | Voluntary: Firm initiated |