Z-2163-2013Class II

Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the...

B Braun Medical Inc / initiated 2012-10-24 / Terminated / root cause: design

B Braun Medical, Inc. began a recall of Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive... on . FDA classified it as Class II and gives the reason as "B. Braun Medical Inc. has become aware of an issue with the Outlook ES Safety Infusion System. The key panel may fail to function as intended if the underlying circuitry fails due to intermittent contact. This issue...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2163-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B Braun Medical Inc
Product
Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.
Product code
FRN Pump, Infusion
Reason for recall
B. Braun Medical Inc. has become aware of an issue with the Outlook ES Safety Infusion System. The key panel may fail to function as intended if the underlying circuitry fails due to intermittent contact. This issue applies specifically to the HOLD button. If this occurs, the Door Processor (DP) will not receive the desired input from the user. This issue can be observed in pumps with key panels
Root cause tag
design
FDA root cause
Component design/selection
Action
B Braun sent an URGENT: MEDICAL DEVICE FIELD CORRECTION letter dated October 24, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Customer Support at 1-800-627-7867 for questions regarding this notice.
Quantity in commerce
28,427
Distribution
Nationwide Distribution.
Product codes and lots
Multiple serial numbers affected. Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2163-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2163-2013Class IITerminatedVoluntary: Firm initiated