Z-2175-2013Class II

Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the...

B Braun Medical Inc / initiated 2012-12-05 / Terminated / root cause: design

B Braun Medical, Inc. began a recall of Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive... on . FDA classified it as Class II and gives the reason as "B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the Outlook Pump Models 620-100 and 620-200. This could have occurred after periods...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2175-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B Braun Medical Inc
Product
Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, and 621-200ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.
Product code
FRN Pump, Infusion
Reason for recall
B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the Outlook Pump Models 620-100 and 620-200. This could have occurred after periods of lost DC (battery) power or out of tolerance low power. This could have led to a delay in therapy, or to delivery of an over or under-dose of a therapy if the user was to rely on
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm, B. Braun Medical Inc., sent an "URGENT: MEDICAL DEVICE FIELD CORRECTION" letter dated December 05, 2012 to its customers. The letter describes the product, problem and actions taken. The firm finished performing "field upgrades" on March 20, 2010 for this issue. Should you require any additional information, please do not hesitate to contact B. Braun Medical Inc., Infusion Systems Customer Service at 1-800-627-7867 (627-PUMP).
Quantity in commerce
14,161
Distribution
Worldwide Disribution: USA (nationwide) and country of: Canada.
Product codes and lots
Models 620-100 and 620-200; multiple serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2175-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2175-2013Class IITerminatedVoluntary: Firm initiated