Z-0763-2013Class I

St. Jude Medical, AGA Medical corporation, AMPLATZER TorqVue FX Delivery System. Model #'s: 9-ITVFX06F45/60...

Abbott Medical / initiated 2013-01-18 / Terminated / root cause: design

St Jude Medical Inc began a recall of St. Jude Medical, AGA Medical corporation, AMPLATZER TorqVue FX Delivery System. Model #'s: 9-ITVFX06F45/60, 9-ITVFX07F45/60, 9-ITVFX007F45/80... on . FDA classified it as Class I and gives the reason as "The distal end of the core wire of the TorqVue FX Delivery System could potentially fracture when exposed to a combination of certain cardiac anatomies and usage conditions". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0763-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
St. Jude Medical, AGA Medical corporation, AMPLATZER TorqVue FX Delivery System. Model #'s: 9-ITVFX06F45/60, 9-ITVFX07F45/60, 9-ITVFX007F45/80, 9-ITVFX08F45/60, 9-ITVFX08F45/80, 9-ITVFX09F45/80, 9-ITVFX10F45/80 9-ITVFX12F45/80, 9-ITVFX13F45/80. Cardiac surgical instrument, used to close atrial septal defects between the right and left atriums.
Product code
DQY Catheter, Percutaneous
Reason for recall
The distal end of the core wire of the TorqVue FX Delivery System could potentially fracture when exposed to a combination of certain cardiac anatomies and usage conditions.
Root cause tag
design
FDA root cause
Device Design
Action
St. Jude Medical sent an "URGENT MEDICAL DEVICE RECALL NOTICE" dated January 18, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were advised to discontinue use of the device and remove it from inventory. Customers should contact their SJM sales representative for questions relating to this recall.
Quantity in commerce
635
Distribution
Nationwide Distribution-including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TX, VA, WA, and WI.
Product codes and lots
all batches

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0763-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0763-2013Class ITerminatedVoluntary: Firm initiated