Z-0931-2013Class II
Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of...
Elekta Limited / initiated 2012-12-19 / Terminated / root cause: software
Elekta, Inc. began a recall of Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process on . FDA classified it as Class II and gives the reason as "The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0931-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Elekta sent an "IMPORTANT FIELD SAFETY NOTICE" dated December 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The IFSN informs the user in the TPS, to give the different volumes different structure names. After importing the patient data set from the TPS, make sure that the structure contours are correct. Users are requested to return the signed acknowledgement. Contact your local Elekta office for questions regarding this notice.
- Quantity in commerce
- 164
- Distribution
- Nationwide Distribution-DC, Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, FL, GA, IL, IA, KY, MD, MA, MI, MN, MS, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, and WI.
- Product codes and lots
- 151055, 151700, 151809, 151695, 151479, 151696, 151867, 151704, 151765, 151860, 151024, 151521, 151803, 152217, 151774, 152220, 151626, 151842, 151847, 151446, 151801, 151953, 151664, 151751, 151352, 151298, 151728, 151504, 151981, 151705, 151977, 152123, 152126, 151955, 105982, 151160, 151634, 151954, 151416, 151074, 151587, 151729, 151694, 151965, 151418, 151674, 151259, 151698, 151699, 152043, 151636, 151811, 151568, 151258, 151584, 151423, 151623, 151627, 151507, 151640, 151933, 151340, 151722, 151810, 151051, 151411, 151360, 151595, 151433, 151611, 152008, 151151, 151714, 151653, 151675, 151538, 152031, 151523, 151519, 151432, 151783, 151410, 151371, 151084, 151387, 151978, 151530, 151328, 151689, 151567, 151598, 151760, 151269, 151557, 151558, 151789, 151398, 151632, 152214, 151301, 151441, 151256, 151391, 151038, 151585, 151327, 151326, 151250, 151130, 151828, 151245, 151672, 105803, 151073, 151956, 152232, 151319, 151959, 151960, 151958, 151435, 151779, 151969, 151848, 151875, 151881, 151882, 151736, 152023, 152080, 152079, 152064, 151734, 151763, 152154, 151110, 151363, 151216, 151173, 151678, 151467, 151961, 151178, 152091, 152051, 152115, 152116, 151708, 152176, 152192, 152098, 152200, 152174, 152246, 152364, 152673, 152425, 151156, 151764, 151886, 151076, 151369, 152175, 105941
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0931-2013 | Class II | Terminated | Voluntary: Firm initiated |