Z-0869-2013Class II

MAYFIELD Composite Series Base Unit Standard, Model (A3101), Rx ONLY, Integra LifeScience Corporation, 4900...

Integra LifeSciences Corporation / initiated 2012-12-21 / Terminated / root cause: design

Integra LifeSciences Corp. began a recall of MAYFIELD Composite Series Base Unit Standard, Model (A3101), Rx ONLY, Integra LifeScience Corporation, 4900 Charlemar Drive, Building A, Cincinnati, OH 45227... on . FDA classified it as Class II and gives the reason as "Integra determined there is a possibility that the Base Units may break during use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0869-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
MAYFIELD Composite Series Base Unit Standard, Model (A3101), Rx ONLY, Integra LifeScience Corporation, 4900 Charlemar Drive, Building A, Cincinnati, OH 45227 The MAYFIELD Base Units are intended to be used to support a patient during diagnostic examination and/or surgical procedures where a rigid support between surgical table and headrest, or skull clamp is necessary, positional freedom is required and where X-ray imaging modalities will be used.
Product code
FWZ Operating Room Accessories Table Tray
Reason for recall
Integra determined there is a possibility that the Base Units may break during use.
Root cause tag
design
FDA root cause
Device Design
Action
Integra LifeSciences Corporation sent an Urgent Medical Device Recall letter dated December 21, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to stop using the affected product immediately and also remove them from service to prevent any further use of the affected product. Integra would provide an alternative base unit to replace those customers could no longer use. Customers were asked to complete the Recall Acknowledgement and Return Form to ensure they received this information. For assistance customers were instructed to call 1-513-533-7944 or 1-513-533-7946
Quantity in commerce
96 units
Distribution
Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA & WI. Internationally to Belgium.
Product codes and lots
Model # A3101 Base Units Serial No's: SB120016, SB120017, SB120019, SB120020, SB120021, SB120022, SB120023, SB120024, SB120025, SB120026, SB120027, SB120028, SB120029, SB120030, SB120031, SB120032, SB120134, SB120036, SB120037, SB120038, SB120040, SB120042, SB120043, SB120044, SB120045, SB120049, SB120050, SB120051, SB120052, SB120053, SB120054, SB120055, SB120056, SB120057, SB120058, SB120059, SB120060, SB120061, SB120062, SB120063, SB120064, SB120065, SB120066, SB120067, SB120068, SB120069, SB120070, SB120071, SB120072, SB120073, SB120074, SB120075, SB120076, SB120077, SB120078, SB120079, SB120080, SB120081, SB120082, SB120083, SB120084, SB120085, SB120086, SB120087, SB120088, SB120089, SB120090, SB120091, SB120092, SB120093, SB120094, SB120095, SB120096, SB120097, SB120098, SB120099, SB120100, SB120107, SB120108, SB120109, SB120111, SB120112, SB120113, SB120114, SB120115, SB120116, SB120117, SB120118, SB120119, SB120120, SB120121, SB120122, SB120129, SB120131, SB120138 & SB120141.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0869-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0869-2013Class IITerminatedVoluntary: Firm initiated