Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1062-2013Class II

Flexible Grip which is part of the Synthes Universal Nail System. Indicated for the conventional...

Synthes GmbH / initiated 2013-01-11 / Terminated / root cause: design

Synthes USA HQ, Inc. began a recall of Flexible Grip which is part of the Synthes Universal Nail System. Indicated for the conventional (non-locking) technique of treating stable diaphyseal fractures on . FDA classified it as Class II and gives the reason as "Synthes is initiating a voluntary recall of the Flexible Grip (Part number 355.28) which is part of the Universal Nail System, due to the potential lack of the ability to thoroughly clean the instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1062-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Flexible Grip which is part of the Synthes Universal Nail System. Indicated for the conventional (non-locking) technique of treating stable diaphyseal fractures.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Synthes is initiating a voluntary recall of the Flexible Grip (Part number 355.28) which is part of the Universal Nail System, due to the potential lack of the ability to thoroughly clean the instrument.
Root cause tag
design
FDA root cause
Device Design
Action
Synthes sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" dated January 11, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter requested consignees examine inventory, remove the recalled products from use and return them to Synthes. If you have any questions, contact the firm at 610-719-5450.
Quantity in commerce
328
Distribution
Nationwide Distribution including the states of AZ, CA, CO, FL, IA, ID, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NV, NY, OH, PA, RI, TN, TX, UT, VT, WA, WI, and WV.
Product codes and lots
Part number 355.28, all lot numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1062-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1062-2013Class IITerminatedVoluntary: Firm initiated