Z-0930-2013Class II

ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical...

Siemens Medical Solutions USA, Inc. / initiated 2013-01-18 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc. began a recall of ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical Solutions USA Inc. business Unit: Ultrasound. The... on . FDA classified it as Class II and gives the reason as "On some ACUSON S2000 ABVS systems, the inside of the transducer pod may contain sharp edges". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0930-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical Solutions USA Inc. business Unit: Ultrasound. The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging of the breast allows volumetric acquisition of the breast image.
Product code
IYO System, Imaging, Pulsed Echo, Ultrasonic
Reason for recall
On some ACUSON S2000 ABVS systems, the inside of the transducer pod may contain sharp edges.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Siemens sent a Customer Safety Advisory notification letter to Siemens Service Personnel on January 18, 2013 and which would be provided to all affected sites during their visit to correct the problem. The notification identified the product, the problem and the action to be taken. The local Customer Service Engineer will perform the test at their site. If the system passes the test, no further action is required. If the ABVS pod fails the inspection, the Service Engineer will correct the sharp edge. For questions regarding this recall call 650-694-5398.
Quantity in commerce
214 affected devices
Distribution
Worldwide Distribution - USA including WV and OH; Internationally to Australia, Austria, Bahrain, Belgium, Brazil, China (P.R.) Estonia, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Macedonia, Malaysia, Mexico, Netherlands, Peru, Philippines, Poland, Portugal, Qatar, Rep of Korea, Romania, Sweden, Taiwan, Thailand, Turkey, Unite Arab Emirates, United Kingdom and Venezuela.
Product codes and lots
Model number 10042541 - AVBS module.  Note: the S2000 system part number is 10041461, however only those systems that include the ABVS module, part number 10042541 are affected. Manufactured prior to 10/30/2012.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0930-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0930-2013Class IITerminatedVoluntary: Firm initiated