Synthes Oracle Spacer System Slap Hammer, Part Number 03.809.690, Lot Number 3723847 The Oracle Spacer is...
Synthes USA HQ, Inc. began a recall of Synthes Oracle Spacer System Slap Hammer, Part Number 03.809.690, Lot Number 3723847 The Oracle Spacer is indicated for use in patients with degenerative disc... on . FDA classified it as Class II and gives the reason as "This recall is being initiated in response to complaints received where it was reported that the proximal end of the Slap Hammer broke during use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2282-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Oracle Spacer System Slap Hammer, Part Number 03.809.690, Lot Number 3723847 The Oracle Spacer is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion combined with supplemental fixation. The Oracle Spacer is indicated as a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace a diseased vertebral body resected or excised during partial and total vertebrectomy procedures or the treatment of tumor or trauma (i.e., fracture), to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of the collapsed vertebral body. The Slap Hammer, 03.809.690, is a dual sided device with two separate attachment options. The Slap Hammer allows a user (surgeon) to remove a previously placed trial implants.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- This recall is being initiated in response to complaints received where it was reported that the proximal end of the Slap Hammer broke during use.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- SYNTHES sent an Urgent Notice Medical Device Recall letter dated January 11, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. The letter requested consignees examine their inventory, remove the recalled product from use and return them to Synthes. Customers were also instructed to completed the attached Verification Section at the end of the letter indicating whether or not they had any of the affected product and the number of devices found. Customers with questions were instructed to call 610-719-5450 or email [email protected]. For questions regarding this recall call 610-719-5000.
- Quantity in commerce
- 27
- Distribution
- Worldwide Distribution - USA including CO, PA, TN, VA, TN, WA, and internationally to Czech Republic.
- Product codes and lots
- Part number 03.809.690, Lot Number 6723847
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2282-2013 | Class II | Terminated | Voluntary: Firm initiated |