Z-0826-2013Class II

ELECTROLASE HYFRECATOR(R) ELECTRODES - SHARP, REF/Catalog Number 7-100-8BX, Disposable Hyfrecator(R) Tips...

ConMed Corporation / initiated 2013-01-28 / Terminated / root cause: manufacturing process

ConMed Corporation began a recall of ELECTROLASE HYFRECATOR(R) ELECTRODES - SHARP, REF/Catalog Number 7-100-8BX, Disposable Hyfrecator(R) Tips, Sterile, Rx Only --- CONMED CORPORATION USA... on . FDA classified it as Class II and gives the reason as "Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0826-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConMed Corporation
Product
ELECTROLASE HYFRECATOR(R) ELECTRODES - SHARP, REF/Catalog Number 7-100-8BX, Disposable Hyfrecator(R) Tips, Sterile, Rx Only --- CONMED CORPORATION USA. Intended to be used as active RF electrodes for desiccation, fulguration and coagulation.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Some devices had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
ConMed sent Urgent Device Recall Letters (dated 1/28/13) and Business Reply Forms to all domestic customers via USPS Priority Mail. International notices were sent via FedEx International Priority to the foreign consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and complete Attachment II response form and return it with unused affected product to: ConMed Corporation 525 French Road, Utica, NY 13502. Customers were asked to return Attachment II even if they don't have any affected product. For questions regarding this recall call 315-624-3533.
Quantity in commerce
Domestic Distribution: 97,450 units; Foreign Distribution: 6,800 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of China, Dubai and Indonesia.
Product codes and lots
Beginning Lot Code 120628X, Ending Lot Code 120917X. [Lot codes on boxes and packaging contain a lot code in the following format: 120917X where 12 represents the year (2012), 09 represents the month (September), 17 represents the day of the month, and X represents the manufacturing shift code (1, 2, 3, 4,5,6 or 9)].

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0826-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0826-2013Class IITerminatedVoluntary: Firm initiated
Z-0827-2013Class IITerminatedVoluntary: Firm initiated